Patented, vibration-insensitive measurements at NIR, SWIR, and MWIR wavelengths.
High performance Twyman-Green laser interferometers offer accurate, repeatable measurement for testing optics at infrared wavelengths
PhaseCam Twyman-Green infrared interferometers accurately measure concave, convex and afocal infrared components, as well as infrared telescopes and lens systems. PhaseCam systems are ideally suited for measuring infrared optics over long paths, for remote mounting in measurement towers or pressure vessels, and for other challenging test setups.
PhaseCam NIR and SWIR Infrared Interferometers
Small and lightweight, PhaseCam NIR and SWIR systems incorporate patented Dynamic Interferometry® technology to acquire data thousands of times faster than a temporal phase shifting interferometer. Because acquisition time is so short, the systems can be used without vibration isolation in the most challenging environments.
PhaseCam NIR and SWIR Specs
Configuration | Twyman-Green interferometer |
Acquisition | Vibration insensitive dynamic or temporal |
Wavelength | NIR (1.064 μm) SWIR (1.55 μm) |
Camera | NIR: 4MP, 12-bit; SWIR: 512 x 512 |
Controls | Motorized focus, beam blocks, contrast |
Software | 4Sight™ Focus included |
Size | NIR: < 39 cm (15.3 in) long SWIR: < 41 cm (15.7 in) long |
Exposure | Min. 30 µsec |
PhaseCam NIR
Operating at 1.064 µm
PhaseCam SWIR
Operating at
1.55 μm
Vibration insensitive
30 microsecond data acquisition time
Reflectivity adjustment
A built-in rotating polarizer helps you achieve maximum fringe contrast and highest-quality data for any sample, from reflectivity <1% up to 100%
Accessory optics
Numerous accessories and options, a fully developed product line.
PhaseCam MWIR Infrared Interferometers
The PhaseCam Twyman-Green MWIR interferometer accurately measures concave, convex and afocal components, as well as optical systems, at mid-IR wavelengths. Its optical system enables measurement of both uncoated optics and high reflectors.
PhaseCam MWIR Specs
Configuration | Twyman-Green interferometer |
Acquisition | Temporal phase-shifting |
Wavelength | 3.39 μm; other wavelengths available |
Camera | 512 x 512 |
Controls | Motorized focus, beam blocks |
Software | 4Sight™ included |
Size | < 85 cm (33 in) long |
Exposure | 500 µsec |
PhaseCam MWIR
Operating at 3.39 µm (other wavelengths available)
4D PhaseCam systems have been deployed for measuring in the most challenging applications: in environmental and vacuum chambers; for long path measurements, without isolation; for with computer-generated holograms for aspheric and freeform optics testing; and as precision alignment aids for complex infrared sensing systems, among many others.
All 4D systems are backed by world-class support every step of the way. Upgrades, accessories and ongoing service will help you get the most of your interferometer investment.
data
analysis
software
Introducing 4Sight Focus
The all-new 4Sight Focus analysis software—included with all 4D interferometers and optical profilers—provides blazing fast acquisition, rich visuals, and a wealth of analysis features.
- Fast acquisition and analysis
- Clean, configurable interface
- Simple to learn
- Deep analysis capabilities
- Easy data output
- Smart integration and automation.
Get In Touch
(520) 294-5600
Location
3280 E Hemisphere Loop, Ste 146
Tucson, AZ 85706
4Dinfo@ontoinnovation.com
Office Hours (Arizona Time)
Mon: 8am – 5pm
Tue: 8am – 5pm
Wed: 8am – 5pm
Thur: 8am – 5pm
Fri: 8am – 5pm
Sat: Closed
Sun: Closed
Send a Message